Evaluation of dose calculation algorithms using different density materials for in-field and out-of-field conditions
DOI:
https://doi.org/10.32471/exp-oncology.2312-8852.vol-41-no-1.12529Keywords:
3D conformal radiotherapy, collapsed cone convolution algorithm, in-field condition, out-of-field condition, stereotactic ablative body radiotherapyAbstract
Summary. Aim: The accuracy of the dose calculation is vital in the stereotactic ablative body radiotherapy (SABR) technique to achieve clinically effective dose distribution for better tumor control. Multiple commercial radiotherapy treatment planning systems (TPS) were implemented with different algorithms, such as Acuros XB in Eclipse and Superposition in XiO. The aim of this study is to investigate five different dose calculation algorithms, namely, pencil beam convolution (PBC), Acuros XB, AAA implemented in an Eclipse system, collapsed cone convolution (CCC) algorithm implemented in Mobius3D and superposition algorithms implemented in the XiO system, and then validate the results against measurements using an Institute of Physical Sciences in Medicine (IPSM) phantom with different density materials for in-field and out-of- field conditions. Material and Methods: The IPSM phantom was used to investigate the dose calculation algorithm performances in four different densities (water, lung, ribs, and dense bone) using different beam configurations, including small beam fields utilised in lung SABR. Five commercial algorithms implemented in two TPS (Eclipse and XiO) and one plan check (M3D) system were used for in-field and out-of-field measurement. Results: In the in-field condition, the Acuros XB algorithm had lower mean differences than the measured dose by the IC ranging from –0.46 to 0.24 for all the densities. In the out-of-field condition, the results of eclipse system: AAA, PBC and Acuros XB algorithms demonstrated underdose point’s measurements by –40% for all densities except for AAA calculations in lung density (overdosed by 40%). The measured points of the superposition algorithms were overestimated to the actual dose less than 30% in water, lung and dense bone. At the same densities, the CCC algorithms showed relatively the lowest differences in percentage compared to the superposition algorithms. Conclusion: Our results showed that the Acuros XB and superposition algorithms are closer to the actual measured dose than AAA, PBC and CCC for majority of the field conditions for water-equivalent, lung, rib and dense bone densities. The CCC algorithm resulted in a better agreement with the measurement of the out-of-field points compared with the other algorithms.
References
ICRU. Determination of absorbed dose in a patient irradiated by beams of X or gamma rays in radiotherapy procedures. Int Commission on Radiation Units and Measurements, Bethesda, MD, 1976: Report No 24.
Fraass B, Smathers J, Deye J. Summary and recommendations of a National Cancer Institute workshop on issues limiting the clinical use of Monte Carlo dose calculation algorithms for megavoltage external beam radiation therapy. Med Phys 2003; 30: 3206–16.
Li JS, Pawlicki T, Deng J, et al. Validation of a Monte Carlo dose calculation tool for radiotherapy treatment planning. Phys Med Biol 2000; 45: 2969.
Vincent W, Teddy K, Cola L, et al. A comparison between anisotropic analytical and multigrid superposition dose calculation algorithms in radiotherapy treatment planning. Med Dosimetry 2013; 38: 209–14.
Van Esch A, Tillikainen L, Pyykkonen J, et al. Testing of the analytical anisotropic algorithm for photon dose calculation. Med Phys 2006; 33: 4130–48.
Hardcastle N, Oborn B, Haworth A. On the use of a convolution-superposition algorithm for plan checking in lung stereotactic body radiation therapy. J Appl Clin Med Physics 2016; 17: 99–110.
Tan YI, Metwaly M, Glegg M, et al. Evaluation of six TPS algorithms in computing entrance and exit doses. J Appl Clin Med Phys 2014; 15: 229–40.
Oyewale S. Dose prediction accuracy of collapsed cone convolution superposition algorithm in a multi-layer inhomogenous phantom. Int J Cancer Ther Oncol 2013; 1: 01016.
Carrasco P, Jornet N, Duch M, et al. Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium. Med Phys 2004; 31: 2899–911.
Davidson S, Ibbott G, Prado K, et al. Accuracy of two heterogeneity dose calculation algorithms for IMRT in treatment plans designed using an anthropomorphic thorax phantom. Med Phys 2007; 34: 1850–7.
Ojala J, Hyödynmaa S, Barańczyk R, et al. Performance of two commercial electron beam algorithms over regions close to the lung-mediastinum interface, against Monte Carlo simulation and point dosimetry in virtual and anthropomorphic phantoms. Phys Med 2014; 30: 147–54.
Thwaites D, Williams JR, Aird EG, et al. A dosimetric intercomparison of megavoltage photon beams in UK radiotherapy centres. Phys Med Biol 1992; 37: 445.
Buzdar S, Afzal M, Todd-Pokropek A. Comparison of pencil beam and collapsed cone algorithms, in radiotherapy treatment planning for 6 and 10 MV photon. J Ayub Med Coll Abbottabad 2010; 22: 152–4.
Fogliata A, Nicolini G, Clivio A, et al. Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm: fundamental characterization in water. Phys Med Biol 2011; 56: 1879.
Bush K, Gagne IM, Zavgorodni S, et al. Dosimetric validation of Acuros® XB with Monte Carlo methods for photon dose calculations. Med Phys 2011; 38: 2208–21.
Stathakis S, Esquivel C, Quino L, et al. Accuracy of the small field dosimetry using the acuros XB dose calculation algorithm within and beyond heterogeneous media for 6 MV photon beams. Int J Med Phys, Clin Engin Radiat Oncol 2012; 1: 78–87.
Childress N, Stephens E, Eklund D, et al. Mobius3D white paper: dose calculation algorithm. Mobius Medical Systems (LP Rev. 0), 2012: 1–10.
Hasenbalg F, Neuenschwander H, Mini R, et al. Collapsed cone and analytical anisotropic algorithm dose calculations compared to VMC++ Monte Carlo simulations in clinical cases. J Phys: Conference Series 2007. IOP Publishing.
Howell RM, Scarboro SB, Kry SF, Yaldo DZ. Accuracy of out-of-field dose calculations by a commercial treatment planning system. Phys Med Biol 2010; 55: 6999.
Kry SF, Followill D, White R, et al. Uncertainty of calculated risk estimates for secondary malignancies after radiotherapy. Int J Radiat Oncol Biol Phys 2007; 68: 1265–71.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2023 Experimental Oncology

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.